Whether you're outfitting a new practice or adding a machine to an existing one, here's exactly what Florida requires before you can put an x-ray system to use — and the one requirement almost everyone underestimates.
If you've just purchased, leased, or otherwise taken possession of an x-ray machine in Florida, there's one rule to keep in mind above all else: the machine must be registered before it's placed into clinical use. There's no 30-day grace period to lean on here — expired or missing registrations are among the most common citations the state issues.
The good news is that the process itself is well-documented once you know what's expected. Here's a complete walkthrough of how registration works with the Florida Department of Health (DOH) Bureau of Radiation Control (BRC) — the agency that oversees more than 64,000 x-ray machines across over 21,000 facilities statewide, from dental and chiropractic offices to hospitals, veterinary clinics, universities, and industrial operations.
Two Separate Requirements — Don't Confuse Them
Before diving into the steps, it's worth flagging a distinction that trips up a lot of new practices:
1. Facility / machine registration — registering the actual x-ray equipment and the location where it's used with the DOH Bureau of Radiation Control. This is what the rest of this guide covers.
2. Operator licensure — a completely separate requirement for the people who take or supervise x-rays, licensed by the Florida Department of Health under its own radiologic technologist program. More on this below.
Who Needs to Register
Under Chapter 64E-5 of the Florida Administrative Code, any facility that operates a radiation machine must register it with the DOH Bureau of Radiation Control before the machine is placed into use. That requirement applies across the board:
- Medical, dental, chiropractic, and podiatric practices
- Veterinary clinics and hospitals
- Hospitals, urgent care centers, and imaging centers
- Universities and academic institutions
- Industrial and analytical x-ray system operators
- Vendors, assemblers, and service companies that install, repair, or sell x-ray machines
The Registration Process, Step by Step
- Gather your information before you start (see the checklist below — it'll save you a second trip through the paperwork).
- If you're building or renovating a room for the machine, get shielding designed and verified by a qualified medical physicist before construction wraps up (see the shielding section below) — retrofitting a room after the fact is far more expensive than building it right the first time.
- Complete DH Form 1107 (Radiation Machine Registration Form) for each facility location.
- Confirm your installer or vendor is registered too — companies that assemble, install, or sell x-ray machines in Florida must hold their own vendor registration on DH Form 1113.
- Get a Report of Assembly for non-certified systems on DH Form 1114, or the federal FDA Form 2579 for certified diagnostic systems (more on this below).
- Submit your registration before the machine is placed into clinical use. This is a hard requirement, not a suggestion.
- Receive your facility registration from the BRC X-Ray Section.
- Keep the department notified. Any change of ownership or change in majority controlling interest must be reported.
What You'll Need on Hand
| Information | Details |
|---|---|
| Entity / facility name | Legal name of the practice, clinic, or company operating the machine |
| Business phone number | Main contact number for the registered entity |
| Physical & mailing address | Where the machine is actually installed and used, and where DOH should send correspondence if different |
| Radiation safety program contact | If you're not already a licensee, DOH requires your radiation safety program to be submitted to the BRC X-Ray Section for review and approval |
| Machine manufacturer & model | As shown on the equipment nameplate |
| Intended use category | Medical, dental, chiropractic, podiatric, veterinary, industrial, or academic |
| Installation date | Confirms your registration is filed before clinical use begins |
| Machine location | Room or suite where the unit physically sits on site |
Shielding & Lead Lining: What Your Walls Actually Need
This is the piece that catches people building out a new room by surprise. Under Rule 64E-5.805, F.A.C., shielding for a new or renovated x-ray room has to be designed or certified by a qualified medical physicist, generally following the methodology in NCRP Report No. 147 for diagnostic x-ray and CT installations (NCRP Report No. 151 for megavoltage/therapy installations). That's a calculation based on your specific room — workload, distance, occupancy of adjacent spaces — not a one-size-fits-all lead thickness you can look up in a table.
What that typically looks like in practice
- Operator/control-booth barriers commonly run around 1.3 mm lead equivalent, though lower-workload dental or mammography-only rooms may only need 0.3–0.5 mm
- Doors are typically built with a single continuous sheet of lead bonded inside, with reinforced hardware to handle the added weight
- Viewing windows use lead glass — lead oxide blended into the glass itself — rather than a lead sheet with a cutout
Don't assume you need lead lining at all: BRC guidance notes that for many general-use rooms, ordinary construction — two thicknesses of drywall on a shared wall, three where the adjoining space isn't under your control — can already provide enough attenuation for scattered radiation. Whether that applies to your room is exactly the calculation a qualified medical physicist is there to run.
One thing worth clarifying: Florida does not require your shielding design to be submitted to the state for review or approval. That responsibility sits with your qualified medical physicist, not DOH. What Florida does require, separately, is that the facility and its x-ray systems be registered with the Department of Health's Bureau of Radiation Control — the process covered throughout this guide.
Once shielding is installed (or a physicist confirms none is needed), a post-construction verification survey confirms it performs as designed. Keep that report on file with your assembly report and physics evaluations — it's one of the first things an inspector asks for if a room's exposure levels are ever in question.
New Equipment: Assembly Reports & Physics Evaluations
Registration gets most of the attention, but two additional requirements kick in the moment a new x-ray unit is assembled and installed — and both are easy to miss if your installer doesn't flag them:
Assembly report — certified diagnostic x-ray systems require a federal FDA Form 2579 (Report of Assembly) filed by the installer; non-certified systems use Florida's own DH Form 1114. Either way, keep a copy on file — it's routinely requested during inspections.
Physics evaluation — a periodic evaluation (generally annual) covering equipment performance testing, radiation dose measurements, safety interlock verification, and image quality assessment. Mammography units require an annual survey by a qualified medical physicist under federal MQSA rules, and any new shielding or facility construction needs a post-construction verification survey, also certified by a qualified medical physicist.
Renewal, Fees & Verification
Florida runs registration on an annual cycle rather than the multi-year cycles you'll see in some other states:
Annual fee, due before October 28
Your annual fee covers both registration and inspection costs, and is due before October 28 each year regardless of when your facility first registered. Missing it is one of the most frequent citations DOH issues, so it's worth calendaring well ahead of the deadline rather than waiting on a reminder notice.
You can look up a facility's registration status and pay fees anytime through the DOH online system at xray.floridahealth.gov, using your registration number and facility zip code.
Questions on state-specific requirements? Contact the Florida Department of Health directly at (850) 245-4888 or RadiationControl@FlHealth.gov.
Don't Forget the Operator
Registering the machine doesn't authorize anyone to use it. Florida licenses operators directly through the Department of Health, across two main tiers:
- General Radiographer, Nuclear Medicine Technologist, or Radiation Therapist — requires completion of a two-year accredited radiologic technology program and passing the ARRT examination (or licensure by endorsement from another jurisdiction's current certification).
- Basic X-Ray Machine Operator (BXMO) — a lower-barrier credential with no formal education program required, though applicants must study the exam guide (or an equivalent) and pass the state exam.
All applicants must be at least 18, of good moral character, pass a background check, and avoid the disqualifying offenses listed in section 435.03, Florida Statutes. Application fees run roughly $40–$45, with exam fees around $150–$200. Licenses renew biennially, requiring 12 hours of continuing education per cycle — including one mandatory HIV/AIDS hour and up to 3 hours of personal development, with the rest focused on technical topics like radiation protection and equipment operation.
Personnel Monitoring: Who Needs a Dosimeter Badge
Personnel monitoring in Florida is triggered whenever a worker is likely to receive more than 10% of an applicable dose limit — for whole-body exposure, that threshold works out to 0.5 rem TEDE per year. In practice:
- Staff who meet the threshold must be issued a whole-body dosimeter badge
- Interventional and hot-lab staff are often issued a ring (extremity) badge in addition to their whole-body badge
- Dose records must be maintained and reviewed on a regular schedule
Don't assume your staff falls below the threshold without checking — your radiation safety program documentation should spell out exactly who's monitored and why.
Inspections & Staying Compliant
DOH conducts more than 15,000 inspections a year statewide, on a schedule that varies by equipment type:
- Mammography and cancer therapy machines: inspected annually
- Dental and podiatry machines: inspected every five years
- Other categories: scheduled based on use frequency
If an inspection turns up a deficiency, you generally have 90 days to correct it. Machines found to pose an immediate threat to public health must be removed from service right away, no grace period. To stay in good standing:
- Pay your annual fee before the October 28 deadline every year
- Report any change of ownership or majority controlling interest promptly
- Keep operator licenses current alongside the machine registration
- Keep assembly reports, physics evaluation results, and shielding survey records on file for inspections
Quick-Reference Checklist
Before you start
- Entity name, phone, physical & mailing address on hand
- Machine manufacturer, model, use category, and install date on hand
- Radiation safety program ready, if you're not already a DOH licensee
- Shielding designed/certified by a qualified medical physicist, if you're building or renovating a room
To register
- Complete DH Form 1107 before the machine is placed into clinical use
- Get your assembly report: FDA Form 2579 (certified systems) or DH Form 1114 (non-certified)
- Confirm your installer/vendor holds their own DH Form 1113 registration
- Get a post-construction shielding verification survey if you built or renovated a room
Ongoing
- Confirm operators hold current Florida DOH radiologic technologist or BXMO licenses
- Issue dosimeter badges to staff who meet the 10% dose threshold and review records regularly
- Complete annual physics evaluations (and annual mammography surveys, where applicable)
- Pay your annual fee before October 28 every year
- Verify status anytime via xray.floridahealth.gov
Helpful Links
- Ionizing Radiation Machines (X-Ray) – Florida DOH
- DH Form 1107 – Radiation Machine Registration Form
- Facility Registration & Payment Portal
- Radiologic Technologist & BXMO Licensing
- License Renewal & Continuing Education
- Chapter 64E-5, F.A.C. – Control of Radiation Hazards
- Rule 64E-5.805, F.A.C. – Shielding and Safety Design Requirements
- FDA Form 2579 – Report of Assembly of a Diagnostic X-Ray System
Bringing a new x-ray system into your practice?
Maven Imaging helps practices across the country source new and refurbished imaging equipment, arrange financing, and get systems installed and running — so registration paperwork is one of the few things left on your plate.
Talk to a Maven Imaging SpecialistThis post summarizes publicly available Florida Department of Health Bureau of Radiation Control information as of July 2026 for general planning purposes and is not legal advice. Requirements, fees, and portal details can change — always confirm current steps and fees directly with the Florida DOH before relying on this guide for a filing.